The pictures are still on my site. I can upload to the form if you would like. Oh, the reason for the picture not linking correctly is the security I enabled in CPANEL to prevent bandwidth theft. If you go [url]http://www.comshyft.com/ESRMeter/[/url] you should be able to view all the files I uploaded I don't plan on taking them down anytime soon.
Front side of board: [URL=http://www.comshyft.com/ESRMeter/front%20side%20of%20board/]http://www.comshyft.com/ESRMeter/front%20side%20of%20board/[/URL]
Back side of...
Leave a comment:
-
Re: Blue ESR Meter "F6" Error
Yes, I did use a DMM. I didn't pay attention to color bands and didn't know that there was a difference between 100R and 100K. But now that its worked out I will pay closer attention in the future. Great learning experience for me. Thanks again everyone.
Leave a comment:
-
Re: Blue ESR Meter "F6" Error
Thank you Radio Fox. I switched those resistors, great find, the unit now tests fine with ".8.8". Calibrated now and all is working. Thank you again.
Leave a comment:
-
Re: Blue ESR Meter "F6" Error
I have both of the POTs in their correct location. The 201 (middle right hand side) and the 103 (middle bottom side), both are in their center position. You should be able to see it in this picture [url]http://www.comshyft.com/ESRMeter/front%20side%20of%20board/PICT0164.JPG[/url].Re: Blue ESR Meter "F6" Error<br /> ...ure [url]http:
Leave a comment:
-
Blue ESR Meter "F6" Error
Hey everyone,
I have been trying to figure out this error. I have not been able to calibrate the unit, not getting anything from the test resistors. Running the self diagnostics give me the F6 error code. I have at this time de-soldered the components checked the resistors, and verified that everything has been put into place. Re-soldered the components listed under the F6 code. Maybe everyone here might be able to stop a misplaced component or point me to test some thing.
I have followed the leads on the board and tested with an ohm meter to see if there are any...